This prospective, multicenter, single-arm trial evaluated the GeminiOne transcatheter edge-to-edge repair system in 120 high-surgical-risk patients with degenerative mitral regurgitation grade ≥3+. The mean age was 72.95 years and 48.3 percent were female. Technical success was 98.33 percent, and 30-day device success was 95.83 percent. At one-year, clinical success was 89.84 percent, freedom from all-cause mortality was 97.48 percent, freedom from major adverse events was 91.61 percent, and freedom from mitral valve reintervention was 98.31 percent. Mitral regurgitation ≤2+ was maintained in 92.24 percent of patients. New York Heart Association (NYHA) class I/II improved from 23.33 percent to 94.74 percent. The authors conclude that GeminiOne showed initial safety and efficacy for high-risk degenerative mitral regurgitation.
You are here:
Novel Transcatheter Edge-to-Edge Repair System for Degenerative Mitral Regurgitation: One-Year Outcomes
Submitted by: Vincent Sier
Source: Journal of the American College of Cardiology: Asia
Keywords: